Keysight E4980A User Manual page 63

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Display Format
MEAS DISPLAY Page
Step 4. Use the softkeys or numeric entry keys to enter the test signal level. When
data is entered with the numeric entry keys, the softkeys change to units
labels (mV, V, uA, mA, A).
Softkey
INCR ++
INCR +
DECR -
DECR --
Table 3-14
Oscillator's voltage level that can be set with INCR ++ / DECR --
INCR ++ / DECR -- (Vrms)
0,
1 m, 2 m, 3 m, •••, 9 m
10 m, 20 m, 30 m, •••, 90 m
100 m, 200 m, 300 m, •••, 900 m
1, 2, 3, •••, 9
10, 20
Table 3-15
Oscillator's current level that can be set with INCR ++ / DECR --
INCR ++ / DECR -- (Arms)
0,
10 , 20 , 30 , •••, 90 
100 , 200 , 300 , •••, 900 
1 m, 2 m, 3 m, •••, 9 m
10 m, 20 m, 30 m, •••, 90 m
100 m
Keysight E4980A/AL Precision LCR Meter
Function
Increases the oscillator's output level with a resolution shown in
Table 3-14
Increases the oscillator's output level with a resolution shown in
Table 3-10
Decreases the oscillator's output level with a resolution shown in
Table 3-10
Decreases the oscillator's output level with a resolution shown in
Table 3-14
1
1
1. The voltage level can be set to more than 2 V only when option 001
is installed.
1
1
1. The current level can be set to more than 20 m only when option
001 is installed.
or
Table
3-15.
to
Table
3-13.
to
Table
3-13.
or
Table
3-15.
63