Display Format
MEAS DISPLAY Page
Table 3-19
DC bias and test signal level-setting limits
Setting value
DC bias
Idc (A)
Idc (A)
Procedure for setting DC bias
Step 1. Press the [Display Format] key.
Step 2. Press the MEAS DISPLAY softkey.
Step 3. Use the cursor keys to select the BIAS field.
Step 4. Use the softkeys or numeric entry keys to enter the test signal level. When
data is entered with the numeric entry keys, the softkey labels are changed to
the appropriate unit labels (mV, V, uA, mA, A).
Softkey
INCR ++
INCR +
DECR -
DECR --
Table 3-20
DC bias signal voltage that can be set with INCR ++ / DECR --
INCR ++ / DECR -- (V)
0,
100 , 200 , 300 , •••, 900
1 m, 2 m, 3 m, •••, 9 m
10 m, 20 m, 30 m, •••, 90 m
100 m, 200 m, 300 m, •••, 900 m
1, 2, 3, •••, 9
10, 20, 30, 40
66
Limit
Test signal
level
Vosc 2 1.15 + Idc 100.2 < 42 V
Vosc(Vrms)
Iosc 2 115 + Idc 100.2 < 42 V
Iosc (Arms)
Function
Increases the oscillator's output level with a resolution shown in
Table 3-20
or
Table
3-21.
Increases the oscillator's output level with a resolution shown in
Table 3-16
to
Table
3-18.
Decreases the oscillator's output level with a resolution shown in
Table 3-16
to
Table
3-18.
Decreases the oscillator's output level with a resolution shown in
Table 3-20
or
Table
3-21.
1
Keysight E4980A/AL Precision LCR Meter