Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
Beep Feature
Functional Description
The beep feature of the E4980A/AL behaves differently depending on which
beep mode is in effect, as shown in
beep modes, see
The beep feature is available in the BIN COUNT DISPLAY page and LIST
SWEEP DISPLAY page.
Table 4-8
Beep output conditions
System beep mode
OFF
ON
To set up the beep output conditions:
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the BEEP field.
Step 4. Set the beep output conditions by pressing the appropriate softkey:
Softkey
FAIL
PASS
Keysight E4980A/AL Precision LCR Meter
Table
"Turning On/Off the Beep Feature" on page
BEEP field of limit
table
FAIL
PASS
FAIL
PASS
Description
Beep is output when the comparator sorting result is OUT OF BIN.
Beep is output when the comparator sorting result is one of BIN1
through BIN9 or AUX BIN.
4-8. For more information on system
155.
Description
Beep is never output.
Beep is output when the comparator
sorting result is OUT OF BIN.
Beep is output when the comparator
sorting result is one of BIN1 through
BIN9 or AUX BIN.
139