Memory:clear; Memory:dim - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
:LIST:VOLTage?
Description
Clears the previous list sweep table, sets AC voltage sweep as the list sweep
parameter, and sets AC voltage points for the AC voltage list sweep. If the list
sweep is not set for AC voltage , the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Range of n
Unit
Example of Use
viVPrintf(AgtE4980A/AL, ":LIST:VOLT 0.5, 0.8, 1, 9.9e37, 1.5"
+ vbLf, 0)
Equivalent key
[Meas Setup] - LIST SETUP - LEVEL[V]

:MEMory:CLEar

Syntax
:MEMory:CLEar DBUF
Description
Clears and disables the data buffer memory. After this command is executed,
measurement data will not be stored in the data buffer memory until execution
of the :FILL command. (No Query)
Equivalent key
[Save/Recall] - SAVE DATA - SAVE & STOP
This key is a combination of both

:MEMory:DIM

Syntax
:MEMory:DIM DBUF,<numeric>
:MEMory:DIM?
Description
Clears the data buffer memory, and sets the data buffer memory's size. An
example of data buffer memory is shown in the chapter on program examples.
Parameter
Range
Preset value
Resolution
Keysight E4980A/AL Precision LCR Meter
<point n>
1 to 201
V
:MEMory:READ?
<Numeric>
1 to 201
201
1
and :MEMory:CLEar.