Display Format
MEAS DISPLAY Page
Displaying Errors instead of Measurement Results
Under certain conditions, the following message will appear instead of the
measurement results. A description of each message is provided below.
Error messages
Message
OVERLOAD
---
INFINITY
Conditions where overload occurs for impedance measurement
Conditions differ depending on the impedance range setting (RANGE field).
Impedance
range
AUTO
HOLD
Table 3-23
Overload range when the impedance range is in the HOLD state for
impedance measurement
Impedance range
( ) (RANGE field)
0.1
1
10
100
300
1 k
Keysight E4980A/AL Precision LCR Meter
Description
This message appears when overload occurs. The returned value is
"9.9E37" when using the SCPI command. Conditions in which
overload occurs differ depending on the impedance range
settings. For details, refer to
This appears when no measurement was performed, or the
measurement result is beyond the display range.
This message appears when the deviation mode of the deviation
measurement function is % and the measurement result cannot be
calculated.
Cond itions
•
E4980A/AL's internal detector detects overload
•
When the DUT's distortion is large
•
When the DUT's response is slow
Conditions differ depending on the impedance range. Refer to
3-23
Measurable range
0 to 0.11
0 to 1.1
0 to 11
0 or more
270 or more
900 or more
Table 3-26 on page
74.
Range in which overload occurs
more than 0.11
more than 1.1
more than 11
n.a.
less than 270
less than 900
Table