Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page

Measurement Functions for the Standard

Functional Description
When you perform load correction, you have to use the reference value of the
standard (either known or pre-measured value). You can select only one of the
available measurement functions as the reference value. For more information
on the available measurement functions, see
page
The load correction feature compensates for error by using a transmission
coefficient derived from the relationship between the standard's reference
value and the actual measurement. Your selected measurement function is
used solely for the calculation of the transmission function.
To set up the measurement function:
Step 1. Press [Meas Setup].
Step 2. Press the CORRECTION softkey.
Step 3. Using the cursor keys, select the FUNC field.
Step 4. Using the cursor keys, select the primary parameter.
Step 5. If there is a secondary parameter and you are presented with another list of
measurement parameters, select one of them by using the softkeys.
For more information on measurement parameters, see
Function" on page
Keysight E4980A/AL Precision LCR Meter
47.
47.
"Measurement Function" on
"Measurement
127