Scanner Interface
How to Use the Scanner Interface Board
Step 17. Use the numeric entry keys or softkeys on the front panel to enter the
measurement point number to perform a LOAD measurement.
Step 18. Connect the standard for test frequency correction of a specified
measurement point number to the measurement contacts of the selected
scanner channel.
Step 19. Use the cursor arrow keys on the front panel to move the cursor to the FREQ
field.
Step 20. Press the MEAS LOAD softkey to measure the standard at a specified
frequency.
Step 21. If necessary, repeat steps 16 to 20 for other measurement points (other
frequencies).
Step 22. Change the scanner's and the E4980A/AL's channel number and repeat steps
4 to 21 until this procedure has been performed for all scanner channels.

Activating the Correction Function

Set the correction function to ON in the following procedure in order to perform
the correction using the correction data.
Step 1. Press the [Meas Setup] - CORRECTION to display the CORRECTION page.
Step 2. Use the cursor arrow keys on the front panel to move the cursor to the SPOT
No. field.
Step 3. Use the numeric entry keys or the softkeys on the front panel to enter the
measurement point number, and confirm that the correction frequencies to
be used (test frequencies of the measurement point number entered) are set
to ON (test frequency and reference values are displayed and OFF is not
displayed).
Step 4. Use the cursor arrow keys on the front panel to move the cursor to the OPEN
field.
Step 5. Press the "ON" softkey to set the OPEN correction function to ON.
Step 6. Use the cursor arrow keys on the front panel to move the cursor to the SHORT
field.
Step 7. Press the "ON" softkey to set the SHORT correction function to ON.
Step 8. Use the cursor arrow keys on the front panel to move the cursor to the LOAD
field.
Step 9. Press the "ON" softkey to set the LOAD correction function to ON.

Correction Data Confirmation

There are two ways to confirm the measurement data for each channel
correction: Use the front panel, and read the data using a query via GPIB.
Keysight E4980A/AL User's Guide