Error Messages
Error Messages
Block data is improper.
-168
Block data not allowed
Block data is not allowed.
C
-140
Character data error
Character data is improper.
-148
Character data not allowed
Character data not allowed for this operation.
-144
Character data too long
Character data is too long (maximum length is 12 characters).
-100
Command Error
Improper command.
41
Correction measurement aborted
Correction data measurement is aborted.
1201
CPU Bd EEPROM write error
An error is generated while writing data to EEPROM. When this error occurs,
contact Keysight Technology's Sales and Service Office or the company from
which you bought the device.
1200
CPU Bd FLASH ROM write error
An error is generated while writing data to FLASH. When this error occurs,
contact Keysight Technology's Sales and Service Office or the company from
which you bought the device.
29
Cycle power of 42841A and Preset
The connection was not established with the 42841A. Cycle the power of the
42841A, then preset the E4980A/AL.
D
90
Data buffer overflow
This error occurs when the amount of data exceeds the defined data buffer
memory size. If this error occurs during the measurement after pressing
[Save/Recall] - SAVE DATA - START LOG, press SAVE & STOP.
-230
Data corrupt or stale
The setting file or correction data to be read cannot be read out with the
current firmware version because the setting file or correction data is corrupt
or old.
-428
Keysight E4980A/AL User's Guide