Handler Interface
Signal Line Definitions
Table E-5
Pin layouts for the list sweep comparator
Pin
Signal name
number
1
/BIN1
2
/BIN2
3
/BIN3
4
/BIN4
5
/BIN5
6
/BIN6
7
/BIN7
8
/BIN8
9
/BIN9
10
/OUT_OF_BINS
11
/AUX_BIN
19-21
N.U.
30
/INDEX
31
/EOM
12-18
22-29
32-36
-464
Signal
Description
d irection
Output
Failure at sweep point 1 (outside limits)
Failure at sweep point 2 (outside limits)
Failure at sweep point 3 (outside limits)
Failure at sweep point 4 (outside limits)
Failure at sweep point 5 (outside limits)
Failure at sweep point 6 (outside limits)
Failure at sweep point 7 (outside limits)
Failure at sweep point 8 (outside limits)
Failure at sweep point 9 (outside limits)
This signal is output when the DUT fails at any of points 1 to 201 during one
sweep cycle.
Output
Failure at sweep point 10 (outside limits)
--
Not used (always off at high level)
Output
SEQ sweep mode: 
This signal is output when the last analog measurement in the sweep cycle is
complete. This means that once the handler has received this signal, the
next DUT can be connected to the UNKNOWN terminal. However,
measurement data are not available until the /EOM signal is output.
STEP sweep mode: 
This signal is output when the analog measurement at each point in the
sweep cycle is complete. This means that once the handler has received this
signal, the instrument can proceed to the next sweep point. However,
measurement data are not available until the /EOM signal is output.
Output
"End of measurement cycle" signal
SEQ sweep mode: 
This signal is output at the end of the sweep cycle. Therefore, when this
signal is output, the measurement data and sorting results are available.
STEP sweep mode: This signal is output when the measurement at each
point is complete (including the comparator test time) and the
corresponding measurement data are available. The comparator test result
is available after the /EOM signal for the last sweep point is output.
Same assignments as for the comparator.