SCPI Command Reference
E4980A/AL commands
Parameter
Preset value
Unit
Equivalent key
[Meas Setup] - CORRECTION - MODE - MULTI
[Meas Setup] - CORRECTION - REF A
[Meas Setup] - CORRECTION - REF B

:CORRection:SPOT[1-201]:OPEN[:EXECute]

Syntax
:CORRection:SPOT[1-201]:OPEN[:EXECute]
Description
Executes OPEN correction at the specified measurement point. (No Query)
Equivalent key
[Meas Setup] - CORRECTION - FREQ - MEAS OPEN

:CORRection:SPOT[1-201]:SHORt[:EXECute]

Syntax
:CORRection:SPOT[1-201]:SHORt[:EXECute]
Description
Executes SHORT correction at the specified measurement point. (No Query)
Equivalent key
[Meas Setup] - CORRECTION - FREQ - MEAS SHORT

:CORRection:SPOT[1-201]:STATe

Syntax
:CORRection:SPOT[1-201]:STATe {ON|OFF|1|0}
:CORRection:SPOT[1-201]:STATe?
Description
Enables the specified measurement point.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - CORRECTION - FREQ - ON
[Meas Setup] - CORRECTION - FREQ - OFF
336
<reference value for primary>, <reference value for
secondary>
0
Depends on
:CORRection:LOAD:TYPE
Description
Enables the specified measurement point
Disables the specified measurement point
Keysight E4980A/AL Precision LCR Meter