Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page
Step 3. Enter the reference value using the softkeys or entry keys. If you use the entry
keys to enter the value, the softkey labels change to unit labels (n, u, m, x1,
1
k).
Softkey
MEASURE
Step 4. Using the cursor keys, select the DEV A field.
Step 5. Select the deviation mode for the primary parameter by pressing the
appropriate softkey:
Softkey
ABS
%
OFF
Step 6. Using the cursor keys, select the (DEV) B field.
Step 7. Select the deviation mode for the primary parameter using the softkeys
described in
1. If the primary parameter of the measurement function is C, the softkey labels show these units: p, n, u, m, x1.
Keysight E4980A/AL Precision LCR Meter
Description
If you want to use a particular device/component as the reference,
connect the DUT and press this key. Then the instrument measures
the DUT once and automatically fills the REF A and (REF) B fields
with the measured values, which now serve as the reference values.
Description
Displays the deviation as the difference from the reference value.
Displays the deviation as a percentage of the reference value.
Turns OFF deviation measurement.
Step
5.
111