Handler Interface

Using the Handler Interface

Using the Handler Interface
Before you can use the handler interface, you need to set up the handler
interface board. Then, set up the limit table for the comparator or the list
sweep table for the list sweep comparator. Finally, enable the handler interface
for signal input/output.
This section provides the handler interface setup procedures for the
comparator and the list sweep comparator.

Setting Up the Handler Interface for Use with the Comparator

To set up the handler interface for the comparator, follow these steps:
Step 1. Press [Meas Setup] - LIMIT TABLE to open the LIMIT TABLE SETUP page.
Step 2. Set up the limit table for bin sorting. For more information, see the chapter on
"Configuring Measurement Conditions."
Step 3. Using the cursor movement keys on the front panel, move the cursor to the
COMP field. The following softkeys appear on the softkey label area.
— ON
— OFF
Step 4. Press the ON softkey. The comparator is enabled.
Step 5. Press the [System] key to open the SYSTEM INFO page.
Step 6. Using the cursor movement keys on the front panel, move the cursor to the
HANDLER I/F field. The following softkeys appear on the softkey label area.
— ON
— OFF
Step 7. Press the ON softkey. The handler interface is enabled for signal
input/output.
Step 8. Press the [Display Format] key to open the MEAS DISPLAY page. Prepare for
measurement by pressing one of the following softkeys: MEAS DISPLAY, BIN
No., and BIN COUNT.
You can also use the BIN No. DISPLAY page to enable or disable the
comparator.
-480
Keysight E4980A/AL User's Guide