Correction:open[:Execute]; Correction:open:state; Correction:short[:Execute]; Correction:short:state - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands

:CORRection:OPEN[:EXECute]

Syntax
:CORRection:OPEN[:EXECute]
Description
Executes OPEN correction based on all frequency points. (No Query)
Equivalent key
[Meas Setup] - CORRECTION - OPEN - MEAS OPEN

:CORRection:OPEN:STATe

Syntax
:CORRection:OPEN:STATe {ON|OFF|1|0}
:CORRection:OPEN:STATe?
Description
Enables OPEN correction.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - CORRECTION - OPEN - ON
[Meas Setup] - CORRECTION - OPEN - OFF

:CORRection:SHORt[:EXECute]

Syntax
:CORRection:SHORt[:EXECute]
Description
Executes SHORT correction based on all frequency points. (No Query)
Equivalent key
[Meas Setup] - CORRECTION - SHORT - MEAS SHORT

:CORRection:SHORt:STATe

Syntax
:CORRection:SHORt:STATe {ON|OFF|1|0}
:CORRection:SHORt:STATe?
Description
Enables SHORT correction.
Parameter
ON or 1
334
Description
Enables OPEN correction.
Disables OPEN correction.
Description
Enables SHORT correction
Keysight E4980A/AL Precision LCR Meter