SCPI Command Reference
E4980A/AL commands
Equivalent key
No equivalent key is available on the front panel.
:MEMory:FILL
Syntax
:MEMory:FILL DBUF
Description
Enables the data buffer memory to store measurement data. After execution of
this command, all measurement data will be stored in the data buffer memory.
If this has already been enabled, this command will do nothing. (No Query)
Equivalent key
[Save/Recall] - SAVE DATA - START LOG
:MEMory:READ?
Syntax
:MEMory:READ? DBUF
Description
Places the data in the data buffer memory into the output buffer. The output
format of this command is the same as the one for FETC:IMP?. In the preset
state, "9.90000E+37,+9.90000E+37,-1,+0" is returned the number of times
specified by :MEMory:DIM. When the data buffer memory is not filled to the
specified size (specified by the
"9.90000E+37,+9.90000E+37,-1,+0" remains for the rest of the data locations.
(Query Only)
Equivalent key
[Save/Recall] - SAVE DATA - SAVE & STOP
This key is a combination of both
:MMEMory:DELete[:REGister]
Syntax
:MMEMory:DELete[:REGister] <numeric>
Description
Deletes the state from the memories. Numbers 0 to 9 are located in the
internal memory, while Nos. 10 to 19 are situated in the external USB memory.
(No Query)
Parameter
Range
Resolution
Equivalent key
[Save/Recall] - No. - DELETE
:MMEMory:LOAD:STATe[:REGister]
Syntax
:MMEMory:LOAD:STATe[:REGister] <numeric>
360
:MEMory:DIM
command),
:MEMory:READ?
<Numeric>
0 to 19
1
Keysight E4980A/AL Precision LCR Meter
and :MEMory:CLEar.