Scanner Interface

Setting Up the Scanner Interface Board

Setting Up the Scanner Interface Board
When you use the scanner interface I/O signals, you must set the pull-up
resistor according to the external DC voltage (EXT. DCV) used with two
switches on the scanner interface board.
The bit switch setting procedure for the pull-up resistor setting is described
below.
Switch setting procedures
Step 1. Disconnect the power cable from the E4980A/AL and allow enough time (a
few minutes) for the internal capacitors to discharge.
Disconnecting the scanner interface board from the unit while power is
supplied or immediately after turning off the power may damage the
scanner interface board and the E4980A/AL
Step 2. Remove the two screws that fix the scanner interface board to the
E4980A/AL's rear panel.
Step 3. Pull off the scanner interface board and disconnect the connected flat cable.
Step 4. Remove the scanner interface board.
Step 5. Set two bit switches (S1, S2) referring to
Table F-11
Setting Bit Switch (S1)
Bit
Defaul t
On
1
1
Off
2
1
Off
3
1
4
On
1. For bits 1 to 3 of bit switch S1, only one bit can be turned on at one time.
-498
Pull-up
EXT. DCV
Resistor
316 
5 V  EXT.DCV  6 V
562 
6 V < EXT.DCV  9 V
9 V < EXT.DCV  15 V
1 k
410 
5 V  EXT.DCV  8 V
(when on)
820 
8 V < EXT.DCV  15 V
(when off)
Table F-11
and
Table
Function
Switch for setting EXT_TRIG
input resistance
Switch for setting /CH_VALID
input resistance