Output:dc:isolation:level:value; Output:dc:isolation[:State]; Output:hpower - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
[Meas Setup] - MEAS SETUP - DCI RNG - HOLD

:OUTPut:DC:ISOLation:LEVel:VALue

Syntax
:OUTPut:DC:ISOLation:LEVel:VALue <numeric>
:OUTPut:DC:ISOLation:LEVel:VALue?
Description
Selects DC isolation range. This command turns the auto range function OFF.
Parameter
Range
Preset value
Unit
Equivalent key
[Meas Setup] - MEAS SETUP - DCI RNG

:OUTPut:DC:ISOLation[:STATe]

Syntax
:OUTPut:DC:ISOLation[:STATe] {ON|OFF|1|0}
:OUTPut:DC:ISOLation[:STATe]?
Description
Enables DC Isolation.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[Meas Setup] - MEAS SETUP - DCI ISO - ON
[Meas Setup] - MEAS SETUP - DCI ISO - OFF

:OUTPut:HPOWer

Syntax
:OUTPut:HPOWer {ON|OFF|1|0}
:OUTPut:HPOWer?
Description
Does nothing. This function is provided because of its command compatibility
with 4284A.
362
<Numeric>
20 |200 |2 m|20 m|100 m
20 m
A
Description
Enables DC Isolation
Disables DC Isolation
Keysight E4980A/AL Precision LCR Meter