Measurement Procedure and Examples
Inductance Measurements
Step 5. Connect the DUT to the test fixture as shown in
Figure 7-22
Connecting DUT
Step 6. Press the [Meas Setup] key to display the MEAS DISPLAY page.
1. Move to the ALC field using the cursor keys to press the ON softkey.
When it is set to ON, the test signal current level through the DUT is set to a
constant value (the value set in the LEVEL field).
Step 7. Press the [Meas Setup] key. Measurements are performed continuously by
the internal trigger, and the measured Ls and Rdc values of the magnetic core
inductor are displayed.
Keysight E4980A/AL Precision LCR Meter