Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page
DC Source
Functional Description
You can set the DC voltage that is output from the DC source output terminal
within the range of -10 V through 10 V.
To set the DC source:
Step 1. Press [Meas Setup].
Step 2. Using the cursor keys, select the DC SRC field.
Step 3. Enter the DC source voltage using the softkeys or entry keys. If you use the
entry keys to enter the value, the softkey labels change to unit labels (mV, V).
Softkey
INCR ++
INCR +
DECR -
DECR --
Table 4-6
Resolution steps when setting the DC source voltage with INCR ++ / INCR --
INCR ++ / INCR -- (V)
0 V
1 m, 2 m, 3 m, •••, 9 m
10 m, 20 m, 30 m, •••, 90 m
100 m, 200 m, 300 m, •••, 900 m
1, 2, 3, •••, 9
10
Keysight E4980A/AL Precision LCR Meter
Description
Increments the DC source voltage in the resolution steps shown in
Table
4-3.
Increments the DC source voltage in steps of 1 mV when lower than
1V or in steps of 10 mV when equal to or higher than 1V.
Decrements Increments the DC source voltage in steps of 1 mV
when lower than 1V or in steps of 10 mV when equal to or higher
than 1V.
Decrements the DC source voltage in the resolution steps shown in
Table
4-3.