SCPI Command Reference
E4980A/AL commands

:COMParator:TOLerance:NOMinal

Syntax
:COMParator:TOLerance:NOMinal <numeric>
:COMParator:TOLerance:NOMinal?
Description
Sets the nominal value for the comparator function's tolerance mode. This can
be set only when the limit mode is set to the tolerance mode.
Parameter
Preset value
Equivalent key
[Meas Setup] - LIMIT TABLE - NOM

:CONTrol:CBIas:STATe

Syntax
:CONTrol:CBIas:STATe {ON|OFF|1|0}
:CONTrol:CBIas:STATe?
Description
Enables the current bias I/F. If the 42841A is connected, the preset status is
ON. This is available only for serial prefix number below:
MY464/SG464 and below for E4980A
MY542/SG542 and below for E4980AL
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
[System] - SYSTEM INFO - CURR BIAS I/F - ON
[System] - SYSTEM INFO - CURR BIAS I/F - OFF

:CONTrol:HANDler:STATe

Syntax
:CONTrol:HANDler:STATe {ON|OFF|1|0}
:CONTrol:HANDler:STATe?
Description
Enables the handler I/F.
330
<Numeric>
0
Description
Enables the current bias I/F
Disables the current bias I/F
Keysight E4980A/AL Precision LCR Meter