Dci Range - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

Configuring Measurement Conditions (Display and Function Related Settings)
MEAS SETUP page

DCI Range

Functional Description
You can set the range for DCI measurement. DCI range can be set to one of five
hold ranges, which are available for DCI measurement.
Before setting the range for DCI measurement, it is necessary to set the DC
bias current isolation feature to ON. For details, refer to
Isolation" on page 97
Model
Standard
When you set the DCI range to HOLD, DCR range and measurement range
are also automatically set to HOLD; when you set the DCI range to AUTO,
DCR range and measurement range are also automatically set to AUTO.
The initial value of DCI measurement range is set to 20 m at HOLD, so
OVERLOAD is displayed when measure the current more than 20 m. In that
case, set the DCI measurement range automatic or 100 m with following
steps.
To set the DCI range:
Step 1. Press [Meas Setup].
Step 2. Using the cursor keys, select the DCI ISO field to set it to ON.
Step 3. Using the cursor keys, select the DCI RNG field.
Step 4. Enter the DCI range using the softkeys or entry keys. If you use the entry keys
to enter the value, the softkey labels change to unit labels (A, mA, A).
Softkey
AUTO
HOLD
INCR +
DECR -
106
Measurement range (A)
20 , 200 , 2 m, 20 m, 100 m
Description
Sets the DCI range to AUTO.
Sets the DCI range to HOLD (manual).
Use this softkey to increment the measurement range. The DCI
range changes to HOLD.
Use this softkey to decrement the measurement range. The DCI
range changes to HOLD.
"DC Bias Current
Keysight E4980A/AL Precision LCR Meter

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4980aalE4980al

Table of Contents