Configuring Measurement Conditions (Display and Function Related Settings)
LIMIT TABLE SETUP Page
Tolerance Mode Nominal Value
Functional Description
When you use tolerance mode as the limit mode for the primary parameter,
you have to configure the nominal value.
The nominal value can be configured even when the primary parameter limit
mode is sequential, but this does not affect sorting.
When you use a negative nominal value in deviation tolerance mode, be
sure to set the lower limit to a value higher than the upper limit because
the lower limit value becomes larger than the upper limit value when they
are converted into absolute values.
To configure the nominal value:
Step 1. Press [Meas Setup].
Step 2. Press the LIMIT TABLE softkey.
Step 3. Using the cursor keys, select the NOM field.
Step 4. Enter the nominal value using the softkeys or entry keys. If you use the entry
keys to enter the value, the softkey labels change to unit labels (n, u, m, x1,
k).
Softkey
INCR ++
INCR +
DECR -
DECR --
Keysight E4980A/AL Precision LCR Meter
Description
Increments the nominal value in steps of your selected number
(1, 2, 5, 10, 20, 50, 100, 200, 500).
Increments the nominal value in steps of 1.
Decrements the nominal value in steps of 1.
Decrements the nominal value in steps of your selected number
(1, 2, 5, 10, 20, 50, 100, 200, 500).
135