Alarms; Errors - Cardinal Health Alaris Directions For Use Manual

Supports guardrails suite mx with guardrails point-of-care software and v9 operating system software. with pc unit, model 8000.
Table of Contents

Advertisement

Alarm
Battery Discharged
Channel Disconnected
Very Low Battery <5 minutes to
system shutdown
Error
Audio System Error
Channel Error
Defective Battery
Alaris
®
System (with v9 Model 8000) DFU
PC Unit Section
Alarms, Errors, Messages (Continued)

Alarms

Operation of all modules stopped
due to insufficient battery charge.
Module(s) disconnected while in
operation or have a communication
problem.
Battery has
of power at current power
consumption rate before operation
stops.

Errors

Main speaker failure.
Error detected. Operation stops on
affected module.
Defective battery.
Meaning
minutes or less
5
Meaning
Response
Connect AC power cord to power
source (alarm silenced). To
continue operation of paused
modules, press
RESTART
affected module.
To silence alarm and clear
message from screen, press
soft key. Reattach
CONFIRM
module, if needed, ensuring it is
securely "clicked" into place at
Module Release Latch. If alarm is
still present, replace module with
an operational instrument.
Connect
power cord to power
AC
source (alarm silenced).
Response
1. Visually check alarm status to
determine whether or not an
operational alarm also needs to
be addressed (red Alarm Status
Indicator lit).
2. Replace PC Unit with an
operational instrument.
To silence alarm and continue
operation of unaffected modules,
press
soft key. Replace
CONFIRM
module with an operational
instrument.
To continue temporary operation,
press
key. Replace
SILENCE
PC Unit with an operational
instrument.
Troubleshooting and Maintenance
key on
1-45