3. Operating Instructions

Carefully read the 1.3 PRECAUTIONS and 1.5 SAFETY TIPS in the 1. IMPORTANT USER INFORMATION section before
attempting to operate and adjust the Cardinal Health™ NPWT PRO family of devices.
WARNING:
The Cardinal Health™ NPWT PRO family of devices should only be used with the supplied A.C. Adapter.
Use of an incorrectly rated adapter could create a shock hazard for the patient or caregiver. The part number for the
adapters can be found in the 6. REPLACEMENT PARTS section of this manual.
3.1 Power ON / OFF
The ON
and OFF
OFF
buttons control the application of power to the device.

3.2 Power-Up Procedure

When turning ON the Cardinal Health™ NPWT PRO family of devices, this is what you should expect to happen.
1. Press the ON
button. All LED indicators will sequentially illuminate during the power-on self-test.
2. Each time the device is turned on, the systems goes through an initialization sequence including the front panel LED
displaying a series of numbers and/or letters.
3. Upon turning ON the device, the dressing should slowly collapse indicating the presence of suction.
4. The Cardinal Health™ NPWT PRO family of devices should be operated at least 22 hours out of every 24 hour period.
Contact your Healthcare Professional if the device is OFF for more than 2 hours in a 24 hour period. Your dressing
must be changed.

3.3 Alarm Operation

Clearing an Alarm Condition
To clear an Alarm Type, remedy the condition using the Alarm Troubleshooting table below. Once the condition is
corrected, the alarm will automatically reset. To manually reset Alarm Type 1-3, turn the therapy device OFF then ON.
The alarm will clear when the power is cycled. Alarm Type 4 cannot be manually reset by cycling power or Muted.
NOTE: Pressing the ON
cannot be Muted or manually reset by cycling power.
NOTE: In the event of an emergency, please contact your treating physician, caregiver, or your local emergency responders.
NOTE: If an Alarm Condition persists and cannot be resolved, please contact Cardinal Health for further assistance.
8
buttons are located on the front top of the control panel. The ON
(MUTE) button after an alarm will silence the alarm for 5 minutes. Alarm condition 4
and