Product Specification

Gas Supply

41–66 psig (nominal 50 psig) clean, dry, medical air and oxygen. External
pressure regulators should be used if source pressures exceed 66 psig.

Pressure Control / Monitoring

Control Range: 2–10 cmH
Monitor Range: 1–11 cmH
<1 or >11 when outside this range.

Oxygen Control / Monitor

Control/Display Range: 21%–100% O
Monitor Range: 17%–104%O
Sensor Type: Long Life Ultrasonic (5+ years)

Flow Monitoring

Range: 0–15 LPM, accuracy ±10%

Power Supply

15 V DC, 3.3 A from included AC adapter (GlobTek, Inc. Model number
GTM21097-5015) or an internal 12 V sealed lead/acid battery. The battery
has a life of four (4) hours from a full charge (up to 16 hours required for
recharge).

Weight

Approx. 15 pounds

Dimensions

8 ½ x 8 ½ x 6 ½ inches (excluding gas inlets, patient outlets, and mounting
brackets)
Pressure Relief (2 systems)
Automatic electronic valve system preset to vent to ambient air when a High
Pressure alarm occurs
Mechanical internal relief valve preset to vent at 3 ±0.15 psig (211 ±11
cmH
O)
2

Flow Limit

Electronic flow limit to prevent the delivery flow from exceeding 3 LPM over
Target Flow or 15 LPM.

Alarm System

Alarms are delayed for 120 seconds after entering the main screen in order
to allow time for patient setup. If the patient setup takes longer than 120
seconds, the alarms can be muted for an additional 120 seconds by
pressing the alarm mute/reset button or the system can be placed into
Standby Mode.
®
AirLife
nCPAP Operator's Manual
O
2
O, accuracy ±1 cmH
2
, accuracy ±3 percentage points
2
accuracy ±3 percentage points
2,
O. Pressure is displayed as
2
3