Setting Up Measurement Conditions and Display

Setting Up Measurement Conditions

Setting Up Measurement Conditions

Selecting measurement parameters

To select the measurement parameters, use the following commands.
You can select one of the following measurement parameter combinations shown in Table
3-1. If you select one parameter using the GPIB command and the resulting parameter
combination is not among those in Table 3-1, the other parameter is automatically changed
to a proper parameter. For example, when the primary parameter is Cp and you select Rs as
the secondary parameter, the primary parameter is automatically changed to Cs.
Table 3-1
Measurement parameter combinations
Each parameter is described below.
Cp:
Cs:
D:
Q:
G:
Rp:
Rs:

Setting up measurement signal (frequency and level)

Setting up frequency
To select the measurement signal frequency (1 kHz/1 MHz), use the following command.
Setting up level
To set the measurement signal level, use the following command.
24
:CALC1:FORM on page 147
:CALC2:FORM on page 150
Primary parameter
Cp
Cs
Capacitance value measured using the parallel equivalent circuit
model
Capacitance value measured using the series equivalent circuit model
Dissipation factor
Quality factor (inverse of D)
Equivalent parallel conductance measured using the parallel
equivalent circuit model
Equivalent parallel resistance measured using the parallel equivalent
circuit model
Equivalent series resistance measured using the series equivalent
circuit model
:SOUR:FREQ on page 187
:SOUR:VOLT on page 188
Secondary parameter
D, Q, G, Rp
D, Q, Rs
Chapter 3