Adc Calibration Status Subsystem - Keysight Technologies M8131A User Manual

16/32 gsa/s digitizer
Hide thumbs Also See for M8131A:
Table of Contents

Advertisement

4
Remote Programming

ADC Calibration Status Subsystem

92
Table 10
Bit Number
0
USB disconnected
The following SCPI commands and queries are supported:
:STATus:QUEStionable:CALibration:ADC[:EVENt]?
:STATus:QUEStionable:CALibration:ADC:CONDition?
:STATus:QUEStionable:CALibration:ADC:ENABle[?]
:STATus:QUEStionable:CALibration:ADC:NTRansition[?]
:STATus:QUEStionable:CALibration:ADC:PTRansition[?]
Table 11
Bit Number
0
Calibration data invalid
1
Calibration necessary
2
Calibration not finished
successfully.
3
Calibration aborted
4
Calibration failed
Upon software start, bit 2 is set and bits 3, 4 are cleared. When a
NOTE
calibration was run, the bits are updated accordingly.
Connection status register
Decimal Value
1
ADC calibration status register
Decimal Value
1
2
4
8
16
Keysight M8131A 16/32 GSa/s Digitizer User's Guide
Definition
USB module connection state
Definition
ADC calibration data is invalid.
ADC re-calibration is necessary.
Cleared after a successful calibration.
ADC calibration was aborted by the user.
ADC calibration failed.

Advertisement

Table of Contents
loading

Table of Contents