Configure Measurement Settings; Select Measurement - Keysight Technologies FieldFox A Series User Manual

Table of Contents

Advertisement

ERTA (Extended Range Transmission Analysis) Mode - Option 209

Configure Measurement Settings

Configure Measurement Settings

On the Receiver unit, select measurement settings such as Frequency range,
Res Be, Points, and so forth. Zeros pan frequency is allowed even though the
frequency is fixed and no stepping is required.
Learn about these settings in

Select Measurement

The following traces are provided for your convenience. They are all configured
as Clear/Write.
Press Meas 1, then choose one or more of the following:
Softkey
Input Power
Output Power
B
Gain or Loss
B/R
Figure 24-2
Input power, Output power, and gain measurement traces
To change the trace numbers of these three measurements, or to select a trace
state other than Clear/Write, press Trace 6.
506
R
"SA Mode Settings" on page
Shown on screen
2
w
r
3
w
B
1
w
G
Keysight N9927-90001 User's Guide
148.
Description
R receiver. Reference power
measured by the Source unit
before entering the DUT.
B receiver. Power out of the
DUT measured at the
Receiver unit.
Gain or Insertion Loss. The B
receiver data divided by the
R receiver data (B/R)
This measurement can be
displayed accurately with, or
without, displaying the R or
B measurements.

Advertisement

Table of Contents
loading

Table of Contents