Configure Subsystem Commands - Keysight Technologies N778 C Series Programming Manual

Polarization instruments
Table of Contents

Advertisement

4
Measurement Operations & Settings

Configure Subsystem Commands

Command:
:CONFigure:MEASurement:SETTing:ACTual?
Syntax:
:CONFigure:MEASurement:SETTing:ACTual?
Description:
Get the index of the setting currently being used.
Parameters:
None
Response:
int
A value >0 is returned if the setting has been stored in FLASH memory (using :CONFigure:MEASurement:SETTing:SAVE), or
has been recalled from FLASH memory (using :CONFigure:MEASurement:SETTing:RECall), and has not been changed since.
0 is returned if the setting has not yet been stored. 0 is returned if the FLASH setting has been deleted
(using:CONFigure:MEASurement:SETTing:ERASe) since the last recall or store. -1 is returned if the setting was changed but
has not been saved yet.
Example:
:conf:meas:sett:act? → +2
Affects:
All instruments
Command:
:CONFigure:MEASurement:SETTing:NUMBer?
Syntax:
:CONFigure:MEASurement:SETTing:NUMBer?
Description:
Get the number of settings. In addition to the settings spaces in FLASH memory, the working memory can hold a setting.
Parameters:
None
Response:
int
Example:
:conf:meas:sett:numb? → +1
Affects:
All instruments
Command:
:CONFigure:MEASurement:SETTing:PRESet
Syntax:
:CONFigure:MEASurement:SETTing:PRESet
Description:
Resets the setting values in the working memory. In contrast to the *RST and System:Preset commands, the previous stored
settings remain in nonvolatile RAM and can be recalled again.
Parameters:
None
52
This section provides the description of the following commands.
Keysight N778xC Series Polarization Instruments Programming Guide