Frequency Response and Isolation Error-Corrections

Removes frequency response of the test setup.
n
Removes isolation in transmission measurements.
n
You can make a response and isolation correction for the following measurements:
reflection measurements
n
transmission measurements
n
combined reflection and transmission measurements
n
Although you can perform a response and isolation correction for reflection measurements,
Hewlett-Packard recommends that you perform an Sll one-port error-correction: it is more
accurate, and it is just as convenient.
1. Press w.
2. Select the type of measurement you want to make.
If you want to make a reflection measurement on PORT 1 (in the forward direction, Sll),
q
leave the instrument default setting.
If you want to make a reflection measurement on PORT 2 (in the reverse direction, S&,
q
press:
3. Set any other measurement parameters that you want for the device measurement: power,
sweep type, number of points, IF bandwidth.
4. lb access the measurement correction menus, press:
5. If your calibration kit is not the 3.5 mm (2.4 mm, HP 8722D) default, press:
If your type of calibration kit is not listed in the displayed menu, refer to the "Modifying
Calibration Kit Standards" procedure, located later in this chapter.
6. 'Ib select a response and isolation correction, press:
7. Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for Srl or PORT 2 for S&.
Include any adapters that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will connect
your device under test.
Optimizing Measurement Results