Perform the a-port Error Corrections
1. Connect adapter A3 to adapter A2 on port 2. (See Figure 5-12.)
Figure 5-12. Two-Port CM Set 1
2. Perform the 2-port error correction using calibration standards appropriate for the
connector type at port 1.
When using adapter removal calibration, you must save calibration sets to the
internal disk, not to internal memory.
3. Save the results to disk. Name the file "PORTI."
4. Connect adapter A3 to adapter Al on port 1. (See Figure 5-13.)
Optimizing Measurement Results 643