Example Program
The following is an example program for performing these same operations over HP-IB:
10 ! File:
20 !
30 ! This demonstrates how to do adapter removal over HP--1B.
40 !
50 ASSIGN
60 !
70 ! Select internal disk.
80 !
90 OUTPUT BNa*"INTD*"
100 !
110 ! Assign file #l to the filename that has a 2-port
120 ! cal previously performed for Port l's connector.
130 !
140 ODTPDT 8Na;"TITFl""FlODCALl"";"
150 !
160 ! Recall the cal set for Port 1.
170 !
180 OUTPUT 6Na;"CALSPORTl;"
190 !
200 !
Assign file #2 to the filename that has a 2-port
210 ! cal previously performed for Port 2's connector.
220 !
230 OUTPUT ONa;"INTD;TITFZ""F20DCAL2"";"
240 !
250 ! Recall the cal set for Port 2.
260 !
270 OUTPUT 6Na;"CALSPORT2;"
280 !
290 ! Set the adapter electrical delay.
300 !
310 ODTPDT ONa;"ADAPl58PS;"
320 !
330 ! Perform the "remove adapter" computation.
340 !
350 !OUTPUT
360 END
,
Optimizing Measurement Results 5 4 7