TOO MANY NESTED SEQUENCES. SEQ ABORTED
Error Number You can only nest sequences to a maximum level of six. The sequence will
abort if you nest more than six.
164
TOO MANY SEGMENTS ORPOINTS
Error Number You can have a maximum of 30 segments or 1632 points in frequency list mode.
In power meter calibrations, you can have a maximum of 12 segments for
50
power sensor cal factors and power loss functions.
TRANSFORM, GATE NOT ALLOWED
Error Number You can perform a time domain transformation only in linear and CW sweep
types.
16
TROUBLE! CHECK SETUP AND START OVER
Service Error Your equipment setup for the adjustment procedure in progress is not correct.
Number 115 Check the setup diagram and instructions
WAITING FOR CLEAN SWEEP
Information
In single sweep mode,
instrument state, if any, have been implemented before taking the sweep. The
Message
command that you have initiated is being processed and will not be complete
until the new sweep is completed. An asterisk * is displayed in the left margin
until a complete fresh sweep has been taken.
WAITING FOR DISK
Information
This message is displayed between the start and finish of a read or write
operation to a disk.
Message