"Modifying Calibration Kit Standards," located later in this section.
This must be done before performing the following sequence.
2.
3. 'lb measure the "TRM THRU," connect the "zero length" transmission line between the two
test ports.
4. 'lb make the necessary four measurements, press:
5. 'Ib measure the "TRM SHORT, " connect the short to PORT 1, and press:
6. Connect the short to PORT 2, and press:
If loads can be connected to both port 1 and port 2 simultaneously,
then the following TRM load measurement can be performed using the
7. lb measure the "TRM LOAD," disconnect the short and connect the TRM load to PORTl.
Refer to "Choosing Calibration Load Standards '
trace settles, press the softkey corresponding to the load used. If a sliding load is
used, press i~~~~~~~ to access the Sliding Load menu. Position the slide and press
9. When all the appropriate load measurements are complete, the load data is measured and
. . . . . . , ; _ ( . t , ; : . , , , J , , . , . , ~ , ~ ~ ~ ~ ~ ~ , ~ ~ ~ , ~ ~ . , . , . . , . . '
11. Repeat the previous TRM load measurement steps for PORT 2.
12. After the measurement is complete, press:
13. 'Ib measure the "ISOLATION" class, press:
Optimizing Measurement Results