CALibration Subsystem

Coarse Adjust the Frequency Reference

:CALibration:FREQuency:REFerence:COARse <setting>
:CALibration:FREQuency:REFerence:COARse?
Allows coarse adjustment of the internal 10 MHz reference oscillator
timebase of the analyzer.
:CALibration:ALL is required after COARse is set.
NOTE
Range:
Front Panel
Access:

Fine Adjust the Frequency Reference

:CALibration:FREQuency:REFerence:FINE <setting>
:CALibration:FREQuency:REFerence:FINE?
Allows fine adjustment of the analyzer internal 10 MHz reference
oscillator timebase.
:CALibration:ALL is required after FINE is set.
NOTE
Range:
Front Panel
Access:

Select the Frequency Corrections

:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Turns on or off the frequency corrections.
Factory Preset
and *RST:
Front Panel
Access:

Align the RF Circuitry

:CALibration:RF
:CALibration:RF?
Performs an alignment of the RF assembly.
5-34
Integer, 0 to 255
System, Alignments, Time Base, Coarse
Integer, 0 to 255
System, Alignments, Time Base, Fine
On
System, Alignments, Freq Correct On Off
Chapter 5