toggles to become view setup when the analyzer is in frequency
offset mode.
specifies the number of the disk volume to be accessed. In
general, all 3.5 inch floppy disks are considered one volume
(volume 0). For hard disk drives, such as the HP 9153A
(Winchester), a switch in the disk drive must be set to defme
the number of volumes on the disk. For more information,
refer to the manual for the individual hard disk drive.
pauses the execution of subsequent sequence commands for x
number of seconds. Terminate this command with (xl.
Entering a 0 in wait x causes the instrument to wait for prior
sequence command activities to linish before allowing the
next command to begin. The wait 0 command only affects
the command immediately following it, and does not affect
commands later in the sequence.
selects the warning annotation for color modification.
brings up the color definition menu. The warning annotation
default color is black.
applies a non-linear phase shift for use with electrical delay
which follows the standard dispersive phase equation for
the active function becomes the WAVEGUIDE CUTOFF
frequency, which is used in the phase equation. Choosing a
Start frequency less than the Cutoff frequency results in phase
errors.
is used to set the amplitude parameter (for example 3 dB) that
defines the start and stop points for a bandwidth search. The
bandwidth search feature analyzes a bandpass or band reject
trace and calculates the center point, bandwidth, and Q (quality
factor) for the specified bandwidth. Bandwidth units are the
units of the current format.
Key Definitions 9-61