Agilent Technologies 8719D User Manual page 448

Network analyzers
Table of Contents

Advertisement

HP 8719D and HP 8720D Measurement Port Specifwations
The following specifications show the residual system performance, including switch
repeatability, after error-correction. The error-correction consists of a fuII 2-port measurement
calibration, including isolation, with an IF bandwidth of 10 Hz and the specified calibration
kit. Calibration temperature is 23O z!c~OC. Measurement must be performed fl0 of calibration
temperature.
HP 8719D/8720D with 3.6 mm Connectors
The following specifications describe the system performance of the HP 8719DB720D network
analyzers. The system hardware includes the following:
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard
Calibration kit: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 85052D
Cables:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 85131F
Load match
a
0.2
0.0
0.05 to 0.5 GHz
0.5 to 2 GHz
42 dB
37 dB
42 dB
-60
-80
-100
I I
I
I
I
I
I I
0.4
06
0.8
10
Frequency Range
2 to 8 GHz
38 dB
37 dB
31 dB
42 dB
33 dB
0.2
Specifications and Measurement Uncertainties
36 dB
28 dB
36 dB
-40
-60
-80
0.6
0.8
0.4