Agilent Technologies 8719D User Manual page 575

Network analyzers
Table of Contents

Advertisement

Error Messages in Alphabetical Order
Error Number This message is displayed if you pressed
217
Then you can enable an auxiliary channel.
ABORTING COPY OUTPUT
Information This message is displayed briefly if you have pressed ILocal) to abort a copy
operation. If the message is not subsequently replaced by error message
Message
number 25, PRINT ABORTED, the copy device may be hung. Press m once
more to exit the abort process and verify the status of the copy device. At this
point, the copy device will probably have an error condition which must be
ADDITIONAL STANDARDS NEEDED
Error Number Error correction for the selected calibration class cannot be computed until you
68
have measured all the necessary standards.
ADDRESSED TO TALK WITH NOTHING TO SAY
Error Number You have sent a read command to the analyzer (such as ENTER 716) without
31
The analyzer has no data in the output queue to satisfy the request.
AIR FLOWRESTRICTED: CHECKFAN FILTER
Error Number Something is restricting the air flow into the analyzer. Check for any debris and
20
clean or replace the fan filter.
ALL REGISTERS HAVE BEEN USED
Error Number You have used all of the available registers; you can store no more instl-ument
200
states even though you may still have sufficient memory. There are 31 registers
available, plus the present instrument state.