Frequency Response and Isolation Error Corrections

You can make a response and isolation correction for the following measurements:
• reflection measurements
• transmission measurements
• combined reflection and transmission measurements
NOTE
Although you can perform a response and isolation correction for reflection measurements,
we recommend that you perform an S
just as convenient.

Response and Isolation Error Correction for Transmission Measurements

This procedure is intended for measurements that have a measurement range of greater than 90 dB.
1. Press
.
Preset
2. Select the type of measurement you want to make.
❏ If you want to make a transmission measurement in the forward direction (S
Meas
Trans: FWD S21 (B/R)
❏ For ES analyzers, if you want to make a transmission measurement in the reverse direction (S
press:
Meas
Trans: REV S12 (A/R)
3. Set any other measurement parameters that you want for the device measurement: power, number of
points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
5. If your calibration kit is different than the kit specified under the
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
page
7-57.
6. To select a response and isolation correction and to start the response portion of the calibration, press:
CALIBRATE MENU
7. Make a "thru" connection between the points where you will connect your device under test.
NOTE
Include any adapters that you will have in the device measurement. That is, connect the
standard device to the particular connector where you will connect your device under test.
Frequency Response and Isolation Error Corrections
11
or on ET models:
(select your type of kit)
RESPONSE & ISOL'N
Calibrating for Increased Measurement Accuracy
one-port error correction; it is more accurate and
TRANSMISSN
CAL KIT [ ]
RETURN
"Modifying Calibration Kits" on
RESPONSE
), press:
21
),
12
softkey, press:
6- 17