calculates and displays the complex ratio of the signal at input
A to the reference signal at input R.
puts the name of the active entry in the display title.
puts the active marker magnitude in the display title.
selects coaxial as the type of port used in adapter removal
calibration.
is used to enter the value of electrical delay of the adapter used
in adapater removal calibration.
provides access to the adapter removal menu.
selects waveguide as the type of port used in adapter removal
calibration.
displays the edit segment menu and adds a new segment to
the end of the list. The new segment is initially a duplicate of
the segment indicated by the pointer > and selected with the
sets the HP-IB address of the analyzer, using the entry controls.
There is no physical address switch to set in the analyzer.
sets the HP-IB address the analyzer will use to communicate
with the external controller.
sets the HP-IB address the analyzer will use to communicate
with an external HP-IB disk drive.
sets the HP-IB address the analyzer will use to communicate
with the power meter used in service routines.
presents a menu for adjusting display intensity, colors, and
accessing save and recall functions for modified LCD color sets
leads to the beginning of the adjustment tests These tests
generate correction constants that are used by the analyzer.
retrieves the full frequency list sweep.
measures only one input per frequency sweep, in order to
reduce spurious signals Thus, this mode optimizes the dynamic
range for all four S-parameter measurements
adds or subtracts an offset in amplitude value. This allows
limits already dellned to be used for testing at a different
response level. For example, if attenuation is added to or
removed from a test setup, the limits can be offset an equal
amount. Use the entry block controls to specify the offset.
displays a dc or low frequency ac auxiliary voltage on the
vertical axis, using the real format. An external signal source
such as a detector or function generator can be connected to
the rear panel AUXILIARY INPUT connector.
selects ASCII format for data storage to disk.
sets the sequence bit in the Event Status Register, which can
be used to generate an SRQ (service request) to the system
controller.