SWEEP TIME TOO FAST
Error Number The fractional-N and digital IF circuits have lost synchronization. Refer to the
12
HP 8719D/ZOD/ZZD Network Anulgzer Service
information.
SWEEP TRIGGER SET TO HOLD
Information
The instrument is in a hold state and, is no longer sweeping. lb take a new
Message
sweep, press LMenu) TRJTGGER MEHU SI$lG*~ or ~~~~~~~~~.
SYNTAX ERROR
Error Number
You have improperly formatted an HP-IB command. Refer to the
33
syntax.
SYST CTRL ORPASS CTRL IN LOCAL MENU
Error Number The analyzer is in talker/listener mode. In this mode, the analyzer cannot
control a peripheral device on the bus. Use the local menu to change to system
36
controller or pass control mode.
TEST ABORTED
Error Number You have prematurely stopped a service test.
113
TEST PORT OVERLOAD, REDUCE POWER
Error Number You have exceeded approximately + 14 dBm at one of the test ports (or 0 dBm
at the A or B sampler, Option 012 Only). When this occurs, reduce the power to
57
a lower level.
THIS LIST FREQ INVALID
Error Number You have set frequencies in the list that are outside of the allowable frequency
133
range of the analyzer. Reduce the frequency range of the list.
. . . . . . . . . . . . . . . . . . . . . . . .