FUNCTION NOT AVAILABLE
Error Number The function you requested over HP-IB is not available on the current
202
instrument.
FUNCTION NOT VALID
Error Number The function you requested is incompatible with the current instnmrent state.
14
FUNCTION NOT VALID DURING MOD SEQUENCE
Error Number You cannot perform sequencing operations while a sequence is being modified.
131
FUNCTION NOT VALID FOR INTERNAL MEMORY
Error Number The function you selected only works with disk files.
201
FUNCTION ONLY VALID DURING MOD SEQUENCE
Error Number You can only use the ;~~~~~~:~~~~~, capability when you are buikiing a
164
sequence. Attempting to use this softkey at any other time returns an error
message and no action is taken.
HP 8720 SOURCE PARAMETERS CHANGED
Error Number Some of the stimulus parameters of the instrument state have been changed,
because you have turned correction on. A calibration set for the current
61
measurement parameter was found and activated. The instrument state was
updated to match the stimuIus parameters of the calibration state.
This message also appears when you have turned on harmonic mode or
frequency offset and the present frequency range cannot be used with one of
these modes.