Frequency Response and Isolation Error Corrections

You can make a response and isolation correction for the following measurements:
• reflection measurements
• transmission measurements
• combined reflection and transmission measurements
Although you can perform a response and isolation correction for reflection
NOTE
measurements, we recommend that you perform an S
correction; it is more accurate and just as convenient.
Response and Isolation Error Correction for Transmission
Measurements
This procedure is intended for measurements that have a measurement range of greater
than 90 dB.
1. Press
.
Preset
2. Select the type of measurement you want to make.
If you want to make a transmission measurement in the forward direction (S
press:
Trans: FWD S21 (B/R)
Meas
For ES analyzers, if you want to make a transmission measurement in the reverse
direction (S
), press:
12
Trans: REV S12 (A/R)
Meas
3. Set any other measurement parameters that you want for the device measurement:
power, number of points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
5. If your calibration kit is different than the kit specified under the
press:
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
Calibration Kits" on page
6. To select a response and isolation correction and to start the response portion of the
calibration, press:
CALIBRATE MENU
or on ET models:
(select your type of kit)
7-57.
RESPONSE & ISOL'N
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
one-port error
11
TRANSMISSN
CAL KIT [ ]
RETURN
RESPONSE
),
21
softkey,
"Modifying
6- 17