Frequency Response (Flatness)

Test Limits

Frequency Range
3 Hz to 3 GHz
> 3 GHz to 6.6 GHz
> 6.6 GHz to 22 Ghz
> 22 GHz to 26.5 GHz
> 26.5 GHz

Test Description

The frequency response test measures the spectrum analyzer's amplitude error as a
function of the tuned frequency. Measurements are made ranging from 50 MHz to the
maximum frequency range of your analyzer. The signal source amplitude is measured with
a power meter to eliminate error due to source flatness. The measured value is normalized
to 50 MHz.
Item
Adapter
Adapter
a
Adapter
Adapter
Cable
Cables
Power Meter
Power Sensor
a
Power Sensor
Power Splitter
Chapter 3
Limit Relative to
50 MHz
±1.5 dB
±2.5 dB
±3.0 dB
±3.5 dB
±3.5 dB
Critical Specifications
(for this test)
Type N(m) to 3.5 mm(m)
Type N(m) to 3.5 mm(f)
2.4 mm(f) to 3.5 mm(m)
3.5 mm(f) to 3.5 mm(f)
3.5 mm, 1 meter
BNC, 120 cm (48 in.)
Compatible with power sensor
Frequency Range: 50 MHz to 26.5 GHz
Frequency Range: 50 MHz to 50 GHz
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
PSA Functional Tests

Frequency Response (Flatness)

Recommended
Agilent Model
1250-1743
1250-1744
11901D
5061-5311
8120-4921
10503A
E4418B
8485A
8487D
11667B
129