CAUTION: CORRECTION OFF: AUXCHANNEL(S) DISABLED
Error Number This message is displayed when correction is forced off due to a stimulus
change that is not compatible with the current calibration while an auxiliary
215
channel is enabled. The auxiliary channels are restored when correction is
turned on by pressing (TZJ ~~~~~~# on OFF .
Error Number There is either a hardware failure in the source or you have attempted to set
179
the power level too high. The analyzer aIlows the output power to be set
higher or lower than the specified available power range. However, these
output powers may be unleveled or unavailable.
level you set is within specifications. If it is, refer to the
Network Analpm Semvice
Error Number Your target value for the marker search function does not exist on the current
data trace.
159
CONTINUOUS SWITCHING NOT ALLOWED
Error Number Your current measurement requires different power ranges on channel 1 and
10
channel 2. lb protect the attenuator from undue mechanical wear, test set hold
will be activated.
In instruments equipped with Option 007, test set hold wilI be activated if the
mechanical transfer switch is required to switch continuously (for example, if
channel 1 is set to measure Sll and channel 2 is set to measure S-22.
The "tsH" (test set hold) indicator in the left margin of the display indicates
that the inactive channel has been put in the sweep hold mode.
COPY: device not responding; copy aborted
Error Number The printer or plotter is not accepting data. Verify the cable connections, HP-IB
addresses, and otherwise ensure that the copy device is ready.
170