Agilent Technologies 8719D User Manual page 584

Network analyzers
Table of Contents

Advertisement

FILE NOT COMPATIBLE WITH INSTRUMENT
Information You cannot recall user graphics that had been saved on an earlier model of
Message
analyzer with a monochrome display. These hles cannot be used with the HP
FILE NOT FOUND
Error Number The requested file was not found on the current disk medium.
192
FILE NOT FOUND OR WRONG TYPE
Error Number During a resave operation, either the file was not found or the type of file was
197
not an instrument state file.
FIRST CHARACTER MUST BE A LETTER
Error Number The ilrst character of a disk file title or an internal save register title must be
42
an alpha character.
FORMAT NOT VALID FOR MEASUREMENT
Error Number Conversion measurements (Z or Y reflection and transmission) are not valid
75
with Smith chart and SWR formats.
FORMATTING DATA
Information The list information is being processed for a list data output to a copy device
and stored in the copy spool buffer. During this time, the analyzer's resources
Message
are dedicated to this task (which takes less than a few seconds.)
FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE
Error Number You can only make frequency offset measurements in the network analyzer
mode.
140
FREQS CANNOT BE CHANGED,TOO MANY POINTS
Error Number The number of points selected for setting the low pass transform frequencies is
204
too high. Reduce the number of points so that the low pass criteria is met.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8720d8722d

Table of Contents