INSUFFICIENT MEMORY, PWR MTR CAL OFF
Error Number There is not enough memory space for the power meter calibration array.
154
Increase the available memory by clearing one or more save/recall registers, or
by reducing the number of points.
INVALID KEY
Error Number You pressed an undefined softkey.
2
Error Number List mode has been turned off in the frequency offset mode because it is
incompatible with your selected LO frequency.
182
LIST TABLE EMPTY
Error Number The frequency list is empty. 'lb implement list frequency mode, add segments
9
to the list table.
LOG SWEEP REQUIRES OCTAVE MINIMUM SPAN
Error Number A logarithmic sweep is only valid if the stop frequency is greater than four
150
times the start frequency. For frequency spans of less than two octaves, the
sweep type automatically reverts to linear sweep.
The frequency domain data points must be harmonically related from dc to the
Information
Message
stop frequency. That is, stop = n x start, where n = number of points. If this
condition is not true when a low pass mode (step or impulse) is selected and
transform is turned on, the analyzer resets the start and stop frequencies. The
stop frequency is set close to the entered stop frequency, and the start
frequency is set equal to stop/n.
MEMORY FOR CURRENT SEQUENCE IS FULL
Error Number All the memory in the sequence you are modifying is filled with instrument
132
commands.