Error Number When you use a sliding load (in a user-defined calibration kit), you must set at
least three slide positions to complete the calibration.
71
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number The ~~~~~~~.. ;3I$$@,J&I$ softkey is not valid unless a calibration is already
69
in progress. Start a new calibration.
NO DISK MEDIUM IN DRIVE
Error Number You have no disk in the current disk unit. Insert a disk, or check the disk unit
41
number stored in the analyzer.
NO FAIL FOUND
Service Error The self-diagnose function of the instrument operates on an internal test
Number 114 failure. At this time, no failure has been detected.
NO FILE(S) FOUND ONDISK
Error Number No files of the type created by an analyzer store operation were found on the
45
disk or the disk drive is empty. If you requested a specific iile title, that file
was not found on the disk.
NO IFFOUND: CHECKR INPUT LEVEL
Error Number The Rrst IF signal was not detected during pretune. Check the front panel R
channel jumper. If there is no visible problem with the jumper, refer to the
5
HP 8719D/2OD/22D Network Armlgm Semrice
NO LIMIT LINES DISPLAYED
Error Number You can turn limit lines on but they cannot be displayed on polar or Smith
144
chart display formats.
1 N-1 5
Error Messafles