Standard Connections for a Response and Isolation Error-Correction for Transmission
If you will be measuring highly reflective devices, such as filters, use the test
device, connected to the reference plane and terminated with a load, for the
isolation standard.
10. To help remove crosstalk noise, set the analyzer as follows:
averages than desired during the device measurement.
11. To measure the calibration standard, press:
12. Return the averaging to the original state of the measurement. For example, reduce the
averaging factor by at least four times or turn averaging off.
13. To compute the isolation error coefficients, press:
The analyzer displays the corrected data trace. The analyzer also shows the notation Cor at
the left of the screen, indicating that the correction is switched on for this channel.
You can save or store the measurement correction to use for later
measurements. Refer to the "Printing, Plotting, and Saving Measurement
Results" chapter for procedures
14. This completes the response and isolation correction for transmission measurements. You
can connect and measure your device under test.
FOR RESPONSE
Figure 5-5.
Measurements
.
.
.
FOR ISOLATION
Optimizing Measurement Results 6-17