Figure 6-57. 'l&t Setup for Continuous Sample Mode

Sample-and-Sweep Mode (One Sweep)

You can use the ~~~~~~~~~ softkey to activate the sample-and-sweep mode. This will correct
the analyzer output power and update the power meter calibration data table during the initial
measurement sweep. In this mode of operation, the analyzer does not require the power meter
for subsequent sweeps. You may use a power splitter or directional coupler, or simply connect
the power sensor directly to the analyzer to measure the power for the initial sweep prior to
connecting and measuring the test device (see Figure 6-58).
The speed of the calibration will be slow while power meter readings are taken (see lhble 6-7).
However, once the sample sweep is finished, subsequent sweeps are power-corrected using
the data table, and sweep speed increases significantly. Once the initial sweep is taken,
sample-and-sweep correction is much faster than continuous sample correction.
If the calibrated power level is changed after the initial measurement sweep is done, the entire
correction table is increased or decreased by that amount and the annotation PCo (indicating
power calibration offset) appears on the display. The resulting power will no longer be as
accurate as the original calibration.
Figure 6-58. Tkst Setup for Sample-and-Sweep Mode
NETWORK ANALYZER
NETWORK ANALYZER
POWER SENSOR
POWER METER