DISK IS WRITE PROTECTED
Error Number The store operation cannot write to a write-protected disk. Slide the
48
write-protect tab over the write-protect opening in order to write data on the
disk.
DISK MEDIUM NOT INITIALIZED
Error Number You must initialize the disk before it can be used.
40
DISK MESSAGE LENGTH ERROR
Error Number The analyzer and the external disk drive aren't communicating properly. Check
190
the HP-IB connection and then try substituting another disk drive to isolate the
problem instrument.
DISK: noton,notconnected, wrong addrs
Error Number The disk cannot be accessed by the analyzer. Verify power to the disk drive,
38
and check the HP-IB connection between the analyzer and the disk drive.
Ensure that the disk drive address recognized by the analyzer matches the
HP-IB address set on the disk drive itself.
Error Number There may be a problem with your disk. Try a new floppy disk. If a new floppy
disk does not eliminate the error, suspect hardware problems.
189
DISK WEAR-REPLACE DISK SOON
Error Number Cumulative use of the disk is approaching the maximum. Copy files as
necessary using an external controller. If no controller is avaiiable, load
49
instrument states from the old disk and store them to a newly initialized disk
using the save/recall features of the analyzer. Discard the old disk.
DOMAIN RESET
Error Number Time domain calculations were reset due to a change in the frequency range,
sweep type, or number of points. Perform a new time domain transform on the
67
new state.
Error Messages 1 O-9